Changes to Step #29
Step Lines
[title] Fans Tested |
[* black] Typical, small tubaxial fans lack sufficient static pressure capability for use in the Fan HALO configuration. We are drawing air in ''through'' a viral filter. Static pressure capability is vital, but we must also balance against filter flow limits. |
[* [* |
[* [* |
[* icon_note] Max CFM rating might seem too high, but remember that is ''without'' back pressure. With resistance of filter and flutter valves, actual flow rate is much lower, about 20 l/min. |
[* black] http://d.digikey.com/HS000xU3ppttIK40NAX00T7 Delta BFB0512HHA-CZMQ has also tested well. It is slightly lower flow, but much quieter. I have worn this while working in the OR for 11 hours. My EtCO2 at 2 hours was 40 with this Delta fan. Its mounting holes must be slightly enlarged to fit 3 mm bolts. |
[* icon_caution] PLEASE DO NOT HORDE FANS. Only obtain the number you will actually ''deploy'' to providers. Hording deprives others in need during this COVID-19 crisis. |
[* black] Other blower fans may yield more ideal combination of power draw, static pressure, flow rate, noise, and availability. [* black] [* icon_note] https://www.digikey.com/product-detail/en/sunon-fans/MF50151VX-B00U-A99/259-1829-ND/7691033 https://www.digikey.com/product-detail/en/sanyo-denki-america-inc/109BC12GC7-1/1688-1035-ND/6191751 https://www.digikey.com/product-detail/en/cui-devices/CBM-5020V-148/102-5585-ND/7623146 |
[* black] Other blower fans may yield more ideal combination of power draw, static pressure, flow rate, noise, and availability. [* black] [* icon_note] https://www.digikey.com/product-detail/en/sunon-fans/MF50151VX-B00U-A99/259-1829-ND/7691033 https://www.digikey.com/product-detail/en/sanyo-denki-america-inc/109BC12GC7-1/1688-1035-ND/6191751 https://www.digikey.com/product-detail/en/cui-devices/CBM-5020V-148/102-5585-ND/7623146 |