[title] Fans Tested
[* black] Typical, small tubaxial fans lack sufficient static pressure capability for use in the Fan HALO configurationBunny PAPR. We are drawing air in ''through'' a viral filter. Static pressure capability is vital, but we must also balance against filter flow limits.
[* black] Typical, small tubaxial fans lack sufficient static pressure capability for use in the Fan HALO configurationBunny PAPR. We are drawing air in ''through'' a viral filter. Static pressure capability is vital, but we must also balance against filter flow limits.
[* black] NMB Technlogies BM5115-04W-B50-L00 blower fan with 3.5 CFM (0.098m³/min), but higher static pressure of 0.803 in H2O (200.0 Pa) provides ample airflow with a single air intake filter and two expiratory flutter valves. https://www.digikey.com/product-detail/en/nmb-technologies-corporation/BM5115-04W-B50-L00/P15682-ND now limited in supply.
[* icon_note] Max CFM rating might seem too high, but remember that is ''without'' back pressure. With resistance of filter and flutter valves, actual flow rate is much lower, about 20 l/min.
[* black] http://d.digikey.com/HS000xU3ppttIK40NAX00T7 Delta BFB0512HHA-CZMQ has also tested well. It is slightly lower flow, but much quieter. I have worn this while working in the OR for 11 hours. My EtCO2 at 2 hours was 40 with this Delta fan. Its mounting holes must be slightly enlarged to fit 3 mm bolts.
[* icon_caution] PLEASE DO NOT HORDE FANS. Only obtain the number you will actually ''deploy'' to providers. Hording deprives others in need during this COVID-19 crisis.
[* black] Other blower fans may yield more ideal combination of power draw, static pressure, flow rate, noise, and availability. 200 to 250 Pa static pressure capability is likely best range.
[* black] If you change fan model, always verify CO2 levels via gas sampling