[title] Inflow/Outflow Balance: Maintain Positive Pressure
|
[* icon_note] I recommend a blower style fan with 200 to 250 static pressure, and viral filters having 3 cm H20 or less flow resistance at 60 LPM. Viral filters with 1.5 cm H20 or lower resistance at 60 LPM are desirable for larger users. |
[* icon_reminder] Fan shall be oriented to pull air in through filter and vent into bag |
[* icon_note] Bag is intentionally NOT stented open. We want the bag to readily collapse. This prevents development of negative pressure during inspiration. |
[* icon_caution] Valves are never instantaneous. Configure your air intake and expiratory fittings to maintain positive to neutral pressure even during inspiration. |
[* icon_caution] Do not massively increase fan power. Keep total liters/min flow to about 20-40 L/min to remain well within anesthesia filter performance envelope.
|